GAO Jintao, ZHEN Zhen, HAN Dan, LIU Bin. Study of Typical Functional Failure and Mechanical Properties of Non-absorbable Suture Anchor[J]. Chinese Journal of Medical Instrumentation, 2023, 47(3): 247-251. DOI: 10.3969/j.issn.1671-7104.2023.03.003
      Citation: GAO Jintao, ZHEN Zhen, HAN Dan, LIU Bin. Study of Typical Functional Failure and Mechanical Properties of Non-absorbable Suture Anchor[J]. Chinese Journal of Medical Instrumentation, 2023, 47(3): 247-251. DOI: 10.3969/j.issn.1671-7104.2023.03.003

      Study of Typical Functional Failure and Mechanical Properties of Non-absorbable Suture Anchor

      • Objective To study the mechanical properties related to the typical functional failure modes of non-absorbable suture anchor in clinical use, and to support product design, development and verification. Methods By retrieving the database of relevant adverse events, the typical functional failure modes of non-absorbable suture anchor were summarized, and the influencing factors of functional failure were further analyzed by studying the mechanical properties related to functional failure. The publicly available test data was retrieved for verification and provided reference for the researchers. Results The typical functional failure modes of non-absorbable suture anchor include anchor failure, suture failure, fix loosening, inserter failure, which are related to the mechanical properties of products, such as screw-in torque and break torque of screw-in anchors, insertion force of knock-in anchors, suture strength, pull-out force before and after system fatigue test and elongation of sutures after fatigue test. Conclusion Enterprises should pay attention to improving the mechanical performance level of products through material, structural design and the suture weaving process to ensure the safety and effectiveness of products.
      • loading

      Catalog

        Turn off MathJax
        Article Contents

        /

        DownLoad:  Full-Size Img  PowerPoint
        Return
        Return