WANG Xiao-yu. Some Problems and Risk Analysis of Infusion Pump Instructions[J]. Chinese Journal of Medical Instrumentation, 2022, 46(3): 323-325. DOI: 10.3969/j.issn.1671-7104.2022.03.018
      Citation: WANG Xiao-yu. Some Problems and Risk Analysis of Infusion Pump Instructions[J]. Chinese Journal of Medical Instrumentation, 2022, 46(3): 323-325. DOI: 10.3969/j.issn.1671-7104.2022.03.018

      Some Problems and Risk Analysis of Infusion Pump Instructions

      • Objective 10 instructions of infusion pumps were contrasted and analyzed in order to explore the problems of instructions, furthermore improve the product safety.Methods Analyzed the scopes of application, requirements of infusion apparatus, maintenance methods and cycles, battery maintenance, and service life, etc.Results There were 5 products in 10 which did not indicate contraindications. One product only provided the thickness range of the tube wall of the infusion apparatus instead of specific brands. The cleaning cycles of 9 products varied from 2 to 12 months. One product did not clarify the service life.Conclusion It is recommended that manufactures should improve the product instructions to ensure the safety of infusion pumps.