YANG Xiao-he, DONG Qin-fang, ZHANG Lan, WANG Yu-hang, ZHU Wen-wu. Discussion on Technical Evaluation Requirements of Allergen Detection Reagent Pre-marketing[J]. Chinese Journal of Medical Instrumentation, 2022, 46(3): 318-322. DOI: 10.3969/j.issn.1671-7104.2022.03.017
      Citation: YANG Xiao-he, DONG Qin-fang, ZHANG Lan, WANG Yu-hang, ZHU Wen-wu. Discussion on Technical Evaluation Requirements of Allergen Detection Reagent Pre-marketing[J]. Chinese Journal of Medical Instrumentation, 2022, 46(3): 318-322. DOI: 10.3969/j.issn.1671-7104.2022.03.017

      Discussion on Technical Evaluation Requirements of Allergen Detection Reagent Pre-marketing

      • In the perspective of technical evaluation, the pre-marketing regulatory requirements of allergen detection reagents in China, America, European Union were compared, and the regulatory risks and performance requirements of this product were analyzed based on the monitoring of post-marketing adverse events, reference standards and domestic and foreign regulatory documents. In view of the "neck-stuck" problems such as the difficulty of clinical trials, the difficulty of finding comparable contrast reagents and the lack of clinical diagnostic gold standards, this paper discusses and gives regulatory suggestions, with a view to providing technical reference for product R&D, production, evaluation, approval and supervision in this field.
      • loading

      Catalog

        Turn off MathJax
        Article Contents

        /

        DownLoad:  Full-Size Img  PowerPoint
        Return
        Return