Zhao Yongjun, Chen Xiao. Versa HD Volume-modulated Accelerator High Voltage Circuit Fault Maintenance[J]. Chinese Journal of Medical Instrumentation, 2020, 44(4): 371-373. DOI: 10.3969/j.issn.1671-7104.2020.04.020
      Citation: Zhao Yongjun, Chen Xiao. Versa HD Volume-modulated Accelerator High Voltage Circuit Fault Maintenance[J]. Chinese Journal of Medical Instrumentation, 2020, 44(4): 371-373. DOI: 10.3969/j.issn.1671-7104.2020.04.020

      Versa HD Volume-modulated Accelerator High Voltage Circuit Fault Maintenance

      • This paper introduces a fast repair methods of Versa HD volume-modulated accelerator's high voltage circuit fault:the key points test method. To identify five key points:①Enter maintenance mode to check for AFC deviation; ② The magnetron waveform MI and PFN waveform PFN V are detected on the maintenance terminal board; ③ Detect thyratron waveform; ④ Check the supply voltage of thyratron; ⑤ HT PSU 600 V DC test 600 V normal or not. The waveform or voltage is also measured to efficiently narrow the fault range to find out the cause of the fault, quickly remove the fault.
      • loading

      Catalog

        Turn off MathJax
        Article Contents

        /

        DownLoad:  Full-Size Img  PowerPoint
        Return
        Return