LIAN Huan, DING Kaizhong, WU Yucheng, CHEN Yonghua, LI Junjun, LI Han. Development of the Extraction Reference Point Beam Diagnostic System for Proton Medical Accelerator[J]. Chinese Journal of Medical Instrumentation, 2019, 43(2): 102-105. DOI: 10.3969/j.issn.1671-7104.2019.02.007
      Citation: LIAN Huan, DING Kaizhong, WU Yucheng, CHEN Yonghua, LI Junjun, LI Han. Development of the Extraction Reference Point Beam Diagnostic System for Proton Medical Accelerator[J]. Chinese Journal of Medical Instrumentation, 2019, 43(2): 102-105. DOI: 10.3969/j.issn.1671-7104.2019.02.007

      Development of the Extraction Reference Point Beam Diagnostic System for Proton Medical Accelerator

      • er to detect the beam quality of the SC200 superconducting cyclotron,measure the beam at the extraction reference and the acceptance of the accelerator is realized.This article mainly introduces the design that use the scintillation screen at the extraction reference to measure the beam profile,position and use the Faraday cup to measure the current intensity with 2.5 level accuracy.The remoted controlling of probes and the acquisition and processing of signal based on LabVIEW and PLC.
      • loading

      Catalog

        Turn off MathJax
        Article Contents

        /

        DownLoad:  Full-Size Img  PowerPoint
        Return