ZHOU Yatong, ZHANG Weiqiang, YU Hongyi. Simulation Study on Radiated Immunity Test of Active Implantable Medical Device[J]. Chinese Journal of Medical Instrumentation, 2019, 43(2): 94-98. DOI: 10.3969/j.issn.1671-7104.2019.02.005
      Citation: ZHOU Yatong, ZHANG Weiqiang, YU Hongyi. Simulation Study on Radiated Immunity Test of Active Implantable Medical Device[J]. Chinese Journal of Medical Instrumentation, 2019, 43(2): 94-98. DOI: 10.3969/j.issn.1671-7104.2019.02.005

      Simulation Study on Radiated Immunity Test of Active Implantable Medical Device

      • In the electromagnetic compatibility standards of active implantable medical devices such as ISO 14117,radiated immunity test above 450 MHz frequency is recommended to be carried out in the electromagnetic shielding room.However,different test locations and the shape/size of the shielding room may lead to very different electromagnetic field distribution in the radiation exposure area of the sample,thus affecting the consistency of the test.With the model built by COMSOL software,this paper analyzes the impact of different parameters,such as size of the room and position of torso simulator on the distribution of field intensity,and reaches results about the distribution of field intensity on the torso simulator area under tow sizes of shielding rooms and two typical test positions.The results show that the experimental consistency of the electric field intensity on the surface directly below the center of the antenna is not good enough,which may affect the repeatability of the test.