WANG Tao, FENG Hansheng, LI Shi, YANG Yang. Design and Implementation of Automated QA Software for Cone Beam CT Image[J]. Chinese Journal of Medical Instrumentation, 2019, 43(1): 25-28. DOI: 10.3969/j.issn.1671-7104.2019.01.007
      Citation: WANG Tao, FENG Hansheng, LI Shi, YANG Yang. Design and Implementation of Automated QA Software for Cone Beam CT Image[J]. Chinese Journal of Medical Instrumentation, 2019, 43(1): 25-28. DOI: 10.3969/j.issn.1671-7104.2019.01.007

      Design and Implementation of Automated QA Software for Cone Beam CT Image

      • In order to improve the speed and accuracy of cone beam CT image quality assurance, an automatic image quality assurance software based on feature extraction is designed and implemented. This paper introduces how the software can realize the positioning of Catphan500 phantoms and the selection of ROI in each module through Canny algorithm and Hough transform circle detection. For the different modules in the Catphan500 phantom, this paper describes the calculation methods of HU accuracy, spatial linearity, HU uniformity and spatial linearity in the software. Finally, the article verifies the feasibility of the software through experiments.
      • loading

      Catalog

        Turn off MathJax
        Article Contents

        /

        DownLoad:  Full-Size Img  PowerPoint
        Return
        Return