Wang Kewu, Xiao Shengxiang, Jiang Lina, Hu Jingkai. Research and Design of a System for Detecting Automated External Defbrillator Performance Parameters[J]. Chinese Journal of Medical Instrumentation, 2017, 41(5): 377-380. DOI: 10.3969/j.issn.1671-7104.2017.05.018
      Citation: Wang Kewu, Xiao Shengxiang, Jiang Lina, Hu Jingkai. Research and Design of a System for Detecting Automated External Defbrillator Performance Parameters[J]. Chinese Journal of Medical Instrumentation, 2017, 41(5): 377-380. DOI: 10.3969/j.issn.1671-7104.2017.05.018

      Research and Design of a System for Detecting Automated External Defbrillator Performance Parameters

      • Objective In order to regularly detect the performance parameters of automated external defibrillator (AED), to make sure it is safe before using the instrument, research and design of a system for detecting automated external defibrillator performance parameters. Methods According to the research of the characteristics of its performance parameters, combing the STM32's stability and high speed with PWM modulation control, the system produces a variety of ECG normal and abnormal signals through the digital sampling methods. Results Completed the design of the hardware and software, formed a prototype. Conclusion This system can accurate detect automated external defibrillator discharge energy, synchronous defibrillation time, charging time and other key performance parameters.