Xu Yongfeng. Microinjection Monitoring System Design Applied to MRI Scanning[J]. Chinese Journal of Medical Instrumentation, 2017, 41(5): 353-356. DOI: 10.3969/j.issn.1671-7104.2017.05.011
      Citation: Xu Yongfeng. Microinjection Monitoring System Design Applied to MRI Scanning[J]. Chinese Journal of Medical Instrumentation, 2017, 41(5): 353-356. DOI: 10.3969/j.issn.1671-7104.2017.05.011

      Microinjection Monitoring System Design Applied to MRI Scanning

      • A microinjection monitoring system applied to the MRI scanning was introduced. The micro camera probe was used to stretch into the main magnet for real-time video injection monitoring of injection tube terminal. The programming based on LabVIEW was created to analysis and process the real-time video information. The feedback signal was used for intelligent controlling of the modified injection pump. The real-time monitoring system can make the best use of injection under the condition that the injection device was away from the sample which inside the magnetic room and unvisible. 9.4 T MRI scanning experiment showed that the system in ultra-high field can work stability and doesn't affect the MRI scans.