      • In order to analyze and study the hemolysis performance of the centrifugal maglev blood pump, which was designed by ourselves, this paper built the mathematical model and computational fluid dynamics analyzed it using Fluent. Then we set up the in vitro hemolysis experiment platform, in case of the design condition, the content of free hemoglobin and hematocrit in plasma were measured in a certain time interval, and calculated the normalized index of hemolysis of the blood pump. The numerical simulation results show the internal static pressure distribution is smooth inside the pump, the wall shear stress inside the pump is less than 150 Pa. Therefore, the red blood cell damage and exposure time is independent. The normalized index of hemolysis is (0.002 9±0.000 7) mg/L, which is in accordance with human physiological requirement.