输液泵质控效率的优化

      Optimization of Quality Control Efficiency of Infusion Pump

      • 摘要: 目的 分析归纳新的质控流程,以提升质控效率。 方法 对2020年的输液泵流速质控数据进行统计分析,总计330次质控数据。 结果 流速合格的泵从第3min开始进入并保持合格的状态,流速不合格的泵从第2min开始进入并保持不合格的状态。 结论 采取第三分钟后进入合格范围并持续2min来判断流速质控合格,这项新规则的准确率高达100%,并且可以将流速质控的时间从60min降低到10min,大大提升了工作效率。

         

        Abstract: Objective Analyze and conclude the new quality control process, so as to improve the quality control efficiency. Methods Statistical analysis was conducted on the flow rate quality control data of infusion pump in 2020, with a total of 330 times of quality control data. Results The pump with qualified flow rate entered and maintained the qualified state from the third minute, while the pump with unqualified flow rate entered and maintained the unqualified state from the second minute. Conclusion Take the third minute to enter the qualified range and continue for 2 minutes to judge that the flow rate quality control is qualified. The accuracy of this new rule is as high as 100%, and the time of flow rate quality control can be reduced from 60 minutes to 10 minutes, which greatly improves the work efficiency.

         

      /

      返回文章