Experimental Study of MAR Algorithm in Metal Artifact Removal of CT Simulator
-
摘要: 目的 探讨MAR算法在CT模拟机金属伪影去除中的应用价值。方法 采用实验扫描带有钛板的CT模体,分别使用常规算法和MAR算法对模体进行扫描。使用伪影指数(AI)、对比噪声比(CNR)和各层面AI值变化情况等客观指标对伪影图像进行统计分析。结果 在伪影指数指标中,MAR算法(10.28±2.60)明显低于常规算法(20.65±5.04);在对比噪声比指标中,MAR算法(7.81±1.12)优于常规算法(5.61±1.36);上述指标在两种算法中均有统计学意义(P<0.01)。受伪影影响的层面中,MAR算法使用前后AI指数下降21.72%~88.40%。结论 MAR算法可以明显减少CT模拟机扫描的金属伪影,同时明显提高CT数据的临床价值。Abstract: Objective To explore the application value of MAR algorithm in metal artifact removal of CT simulator. Methods CT phantom with titanium plate was scanned using conventional algorithms and MAR algorithms, respectively. Artifact index(AI), contrast-to-noise ratio(CNR) and AI values at different slices were used to analyze the artifact images. Results In artifact index, MAR algorithm (10.28±2.60) is significantly lower than conventional algorithm (20.65±5.04); In contrast-to-noise ratio index, MAR algorithm (7.81±1.12) is better than conventional algorithm (5.61±1.36). The above indicators were statistically significant in both algorithms (P<0.01). In the slices affected by metal artifacts, the artifact index decreased by 21.72%~88.40% after the MAR algorithm. Conclusion MAR algorithm can significantly reduce the metal artifacts and improve the clinical value of CT data.
-
Keywords:
- MAR algorithm /
- CT simulator /
- metal artifacts /
- phantom
-
-
[1] MURAZAKI H, FUKUNAGA J, HIROSE T, et al. Dosimetric assessment of a single-energy metal artifact reduction algorithm for computed tomography images in radiation therapy[J]. Radiol Phys Technol, 2019:1-9.
[2] KATSURA M, SATO J, AKAHANE M, et al. Current and novel techniques for metal artifact reduction at CT:practical guide for radiologists[J]. Radio Graphics, 2018, 38(2):450-461.
[3] MAHNKEN A H, RAUPACH R WILDBERGER J E, et al. A new algorithm for metal artifact reduction in computed tomography:in vitro and in vivo evaluation after total hip replacement[J]. Invest Radiol, 2003, 38(12):769-775.
[4] KRIMI S, PAMELA C, CHRISTOPH W, et al. Segmentation of artifacts and anatomy in CT metal artifact reduction[J]. Med Phys, 2012, 39(10):5857-5868.
[5] WANG J. Iterative image reconstruction algorithms for CT metal artifact reduction:a review[J]. Recent Patents Med Imaging, 2013, 3:111-117.
[6] 刘卓,魏拓,洪楠,等.双能量CT联合金属伪影削减算法抑制股骨柄假体金属伪影的模型研究[J].放射学实践, 2019, 34(3):332-336. [7] 通用电气公司.获得MAR图像的方法及装置:中国, CN201610655886.2[P]. 2018-02-23. [8] 高维青,陈毅,王晶波,等.能谱锥形束CT减少口腔金属伪影的效果评价[J].中国口腔颌面外科杂志, 2018, 16(6):48-52. [9] 吴坚,徐冶敏,韩永顺,等. CT能谱成像技术去除眼眶金属植入物伪影的临床应用[J].组织工程与重建外科杂志, 2019, 15(1):17-19. [10] KATSUMURA S, SATO K, IKAWA T, et al. "High-precision, reconstructed 3D model" of skull scanned by conebeam CT:Reproducibility verified using CAD/CAM data[J]. Leg Med, 2015(11):37-43.
[11] TANG H, YANG D, GUO S, et al. Digital tomosynthesis with metal artifact reduction for assessing cementless hip arthroplasty:a diagnostic cohort study of 48 patients[J]. Skeletal Radiol, 2016, 45(11):1523-1532.
[12] FRAGA-MANTEIGA E, SHAW D J, DENNISON S, et al. An optimized computed tomography protocol for metallic gunshot head trauma in a seal model[J]. Veterin Radiol Ultrasound, 2014, 55(4):393-398.
[13] 胡永志,梁媛. Revolution宝石能谱CT在减少腰椎内固定引起的硬化伪影中的应用[J].中外医疗, 2019, 38(1):181-184. [14] 颜森森,罗敏,汪奇慧,等.单源能谱CT在消除金属内固定伪影中的应用[J].医学影像学杂志, 2016, 26(1):94-95, 98. [15] 房加高,邹月芬,徐海,等.多排螺旋CT扫描MAR算法在去除膝关节金属伪影中的应用价值[J].中外医疗, 2018, 37(28):10-12. -
期刊类型引用(6)
1. 唐堂,魏佳琦,王慧婷. 金属去伪影算法MAR在青少年脊柱侧弯术后能谱CT扫描中减少金属伪影的效果. 影像研究与医学应用. 2024(23): 173-175 . 百度学术
2. 胡小洋,韩磊,汪洋,孙磊,倪春霞,刘颖. TTF电极阵列及头皮保护技术对放疗剂量分布影响的初步研究. 中国医学物理学杂志. 2022(02): 169-176 . 百度学术
3. 赵鹏,苏秀军,王宇,孔得福,杨兆禄. 医用射波刀系统的质量保证测试研究. 世界复合医学. 2022(01): 73-77 . 百度学术
4. 谭蕾,李克敏,黄慧琼,贾凤林,宁刚. O-MAR联合iDose~4技术在改善宫颈癌后装治疗CT定位图像金属施源器伪影中的应用. 肿瘤预防与治疗. 2022(06): 544-548 . 百度学术
5. 劳峥,毕帆,樊文慧,胡海生,涂文勇,石慧烽. 去伪影算法对头颈部肿瘤患者钛板植入术后放疗的剂量影响. 中国口腔颌面外科杂志. 2021(03): 248-252 . 百度学术
6. 景生华,李傲梅,季小芹,沈泽天,周含,李兵,朱锡旭. 金标伪影对射波刀剂量计算及分布的影响. 中国医学物理学杂志. 2020(08): 996-999 . 百度学术
其他类型引用(0)
计量
- 文章访问数: 30
- HTML全文浏览量: 3
- PDF下载量: 8
- 被引次数: 6