软式内镜洗消设备的优化改造

      Optimization of Soft Endoscope Cleaning and Disinfection Equipment

      • 摘要: 该文根据国家相关软式内镜清洗标准新版本的要求,按照与前版本标准的差异,对原有洗消设备进行升级改造。比较改造前后清洗单条内镜的平均清洗时间,统计医院一年以来的感控监测结果发现,单条内镜的平均清洗时间从改造前的20 min 55 s提升到改造后的17 min 16 s。感控检测数据中合格率一项,没有明显差别。检出的菌落数中,改造前为6.8,改造后为3.1。通过改造洗消设备,不仅符合最新版本标准要求,而且提高了电子内镜消毒效率和效果。

         

        Abstract: According to the national soft endoscope cleaning standard and the difference between the previous and latest version, the origin decontamination equipment is upgraded. Sensory control monitoring results in the last year shows that the average cleaning time for a single endoscope has been reduced from 20 min 55 s to 17 min 16 s. The number of bacterial colony has been reduced from 6.8 to 3.1. The pass rate of sensory control monitoring keeps steady. The upgraded decontamination equipment not only meets the requirements of the latest version of the standard,but also improves the efficiency and effectiveness of electronic endoscope disinfection.

         

      /