Abstract:
Microwave thermal ablation technology is widely used in the treatment of liver tumors because of its minimal invasiveness and small side effects. The precise control of the thermal dose largely determines the therapeutic effect of microwave thermal ablation. However, the current magnetron-based microwave thermal ablation device has the disadvantages of poor power output stability and high operating voltage. In view of the above problems, this paper selected the microwave solid-state source as the core device of microwave output, and designed a dual-frequency microwave thermal ablation system based on 2 450 MHz and 433 MHz. The system used the power detection circuit to perform PID feedback control on the actual output power of the system. The experimental results show that the maximum output power of the system is less than 1 W. The PID algorithm further enhances the accuracy and stability of the system output power while improving the system security. The dual-frequency microwave thermal ablation system designed in this paper can provide a safe and reliable experimental platform for subsequent research.