基于胸部电阻抗法的无拘束心排血量检测系统

      Unconstrained Cardiac Output Measurement System Based on Thoracic Electrical Impedance Method

      • 摘要: 该文开发了基于胸部电阻抗法和嵌入式技术的无拘束心排血量检测系统。利用PIC嵌入式单片机PIC18F4520的PWM功能以及滤波电路产生频率为50 kHz、有效值为2 mArms的恒流正弦信号,并作为激励信号施加在位于两耳后部和右下腹部的激励电极上;通过位于胸部正中线上的检测电极提取电信号并经检测电路处理后得到人体胸部的电阻抗,据此计算心排血量。采用基于染料指示剂稀释法的检测装置和本研究开发的基于胸部电阻抗法的无拘束心排血量检测系统进行了对比实验。实验结果表明,两种方法具有较好的一致性。

         

        Abstract: An unconstrained cardiac output detection system based on thoracic electrical impedance method and embedded technology is developed in this study. PWM function of PIC embedded MCU, PIC18F4520, and filter circuit are used to generate a constant current sinusoidal signal with a frequency of 50 kHz and a current RMS of 2 mArms, which is applied as an exciting signal to the exciting electrodes located on the positions behind ears and on right lower abdomen. The thoracic electrical impedance is measured by the detecting electrodes pasted along the medial line on the front of the thorax, and is used to calculate cardiac output. A comparative experiment with the thoracic impedance method and the dye dilution method is carried out and the results are investigated. The experimental results show that the two methods have good consistency.

         

      /