一种用于植入式神经刺激器测试的体模设计

      A Torso Simulator Design for Implantable Nerve Stimulator Test

      • 摘要: 该文针对植入式神经刺激器的最新版国际标准ISO 14708-3:2017,特别是有关射频电磁干扰防护的新要求做了介绍。为满足新要求,该文提出了一种用于植入式神经刺激器测试的体模设计。该设计包括相应的容积导体、电极以及网格托架,可模拟射频电磁干扰防护测试中植入式神经刺激器的实际运行环境。该文最后部分对体模进行了性能验证。

         

        Abstract: This paper introduces ISO 14708-3:2017, the new edition of the international standard for implantable neurostimulator, and emphasizes the new requirements in the clause of protection from RF electromagnetic interference. To meet this new requirements, this paper presents a design of torso simulator for the testing of implantable neurostimulator. The design includes volume conductor, electrodes and grids, which can simulate the actual operating environment of implantable neurostimulator in RF electromagnetic interference testing. The torso simulator is verified by performance in the last part of the paper.

         

      /