基于CC2530单片机的心电监测分析仪设计

      Design of ECG Monitoring Analyzer Based on MicrocontrollerCC2530

      • 摘要: 目的以CC2530单片机为核心,设计出一个远程无线心电监测系统。方法在CC2530单片机上实现心电信号的采集转换和数据处理,然后通过ZigBee将数据传输到协调器,实现了心电信号和心率变异性数据的实时监测的功能。结果经测试,该系统的最大误差为3 beat/min,平均误差为1.6 beat/min。符合中华人民共和国医药行业标准。结论该设备有便携测量精度高的特点,具备一定参考价值和实用性。

         

        Abstract: Objective A remote wireless electrocardiogram (ECG) monitoring system is designed by using the CC2530 micro-controller as the device core. Methods Acquisition, conversion and data processing for ECG signals are realized on CC2530 micro-controllers. And the ECG data is transmitted to the coordinator by using ZigBee. It realizes the real-time monitoring of ECG signals and heart rate variability (HRV) data. Results The test results show that the maximum error of the designed ECG monitoring analyzer is 3 beats per minute and the average error is 1.6 beats per minute, which can meet the requirement of pharmaceutical industry standards of the People's Republic of China. Conclusion The ECG monitoring analyzer has good portability, high measurement precision and good practical application values.

         

      /