一种桡动脉触力传感器的检测装置

      A Testing Device for Pulse Graph Force Transducer Measured over Radial Artery

      • 摘要: 桡动脉触力传感器技术是中医脉象诊断的共性关键技术,随着国际标准脉搏波触力传感器(ISO 19614)发布实施,如何对桡动脉触力传感器进行检测的需求也将浮出水面。该文根据中医脉象采集原理和特点,结合国际标准要求,研制了一种将桡动脉触力分解成动态力和静态力分别进行检测,然后进行合成的检测装置,以降低大量程、高精度标准传感器制作的技术难度和制造成本。经对检测装置进行验证,其结果符合国际标准规定的相关要求。

         

        Abstract: Pulse graph force transducer measured over radial artery is a common and key technology of pulse diagnosis in Traditional Chinese Medicine (TCM). In order to solve the problem to conduct the technical test of transducer with the publication of international standard(ISO 19614), a test device has been developed based on principles and characteristics of pulse acquisition in TCM and international standard requirement. The device is a synthetic device designed to test both dynamic and static force to reduce the cost and complexity. The results meet the relevant requirements of international standards through verification.

         

      /