过氧化氢等离子灭菌技术对医用材料细胞毒性影响的评价

      Evaluation of in Vitro Cytotoxicity Tests for Medical Material Using Hydrogen Peroxide Gas Plasma Sterilization

      • 摘要: 该文就过氧化氢等离子灭菌技术对11种医用材料的细胞毒性进行探讨,并与传统常规的环氧乙烷灭菌方法比较。结果显示11种材料中仅尼龙材料不适宜采用过氧化氢等离子灭菌。

         

        Abstract: This paper aims to discuss the evaluation of in vitro cytotoxicity tests for eleven different medical materials using hydrogen peroxide gas plasma sterilization. Compared with ethylene oxide sterilization, the results show that only nylon material is not suitable for using hydrogen peroxide gas plasma sterilization.

         

      /