上海市二级乙等医院CT机质量控制检测

      CT Quality Control Test from the Second Level 2nd Class Hospitals in Shanghai

      • 摘要: 该研究目的是评价上海市二级乙等公立医院和民营医院CT机总体应用质量状况。检测和评价方法按照GB 17589—2011《X射线计算机断层摄影装置质量保证检测规范》的要求进行。对59家医院的62台CT机进行了质量控制检测,高对比分辨力的平均值为6.8 lp/cm,低对比可探测能力的平均值为2.4,CT值(水)的平均值为3.0 HU,噪声的平均值为0.43%,均匀性的平均值为2.3 HU。有20台CT机的5个检测项目均合格,合格率仅为32%。设备老化和缺乏预防性维护保养是导致CT影像质量下降的主要原因。

         

        Abstract: The purpose of this study is to assess the present situation about the overall performance quality of CT scanners from Second Level 2nd Class hospitals and private hospitals in Shanghai. The test and evaluation methods refer to Specification for Quality Assurance Test for Computed Tomography X-ray Scanners (GB 17589-2011). A total of 62 CT scanners from 59 hospitals were tested. The mean value of high-contrast resolution was 6.8 lp/cm, low contrast detectability was 2.4, CT number of water was 3.0 HU, noise was 0.43% and uniformity was 2.3 HU. In quality control tests, 20 of 62 CT scanners were qualified, and the percent of pass was only 32%. The equipment aging and lack of preventive maintenance were main reasons for performance degradation of CT scanners.

         

      /