EDOSE进行调强剂量验证结果的评估

      Evaluation of the EDOSE for IMRT Delivery Quality Assurance

      • 摘要: 目的 使用EDOSE剂量验证系统基于加速器电子射野影像系统(EPID)对静态调强放疗(IMRT)计划剂量验证结果的评估。方法 分析不同射野下EDOSE对EPID成像板剂量响应均匀性的修正;不同大小射野gamma通过率;比较EDOSE反推计算的相关参数与计划系统的差异。结果 修正后EPID不同射野均匀区域图像灰度值的均匀性较好;γ(3mm/3%)分析不同射野的通过率均在理想范围。计划系统和EDOSE剂量验证系统分别计算危及器官的体积并比较,仅右肺存在统计学差异;计划系统和EDOSE剂量验证系统反推计算DVH比较,部分结果存在统计学差异。结论 EDOSE与计划系统具有契合度较高的计算和分析结果。

         

        Abstract: Objective Evaluation of intensity-modulated radiation therapy(IMRT) delivery quality assurance(DQA) using EDOSE software based on accelerator electronic portal imaging device (EPID). Methods Analysed the dose response uniformity correction of EPID imaging plate in different fields and the gamma pass rates about the fields; And compared the calculated parameters of the EDOSE and the Pinnacle planning station. Results The uniformities on the homogeneous region of the fields are good after who corrected by EDOSE. The gamma(γ, 3mm/3%) pass rates are in the ideal range. The volume of the organs at risk(OAR) have no statistical significance except the right lung; A part of the parameters have statistical significance. Conclusions The results of the calculation and analysis of EDOSE has a high agreement with the planning system.

         

      /