不同扫描条件下CBCT影像质量及其HU值长期稳定性分析

      Evaluation of Cone Beam CT Image Quality and Analysis of Its Long-term Stability on HU Values under Different Conditions

      • 摘要: 目的 分析研究不同扫描条件下CBCT影像质量及其HU值长期稳定性。方法 利用OBI系统对Catphan 504模体在不同临床条件下进行扫描,利用方差分析近一年影像数据的差异性。结果 不同条件下CBCT影像密度分辨力均满足Varian验收要求;Full-fan扫描条件下空间分辨力(≥7 lp/cm)优于Half-fan(≥5 lp/cm);空间线性距离及角度显示精度均很好(s<±1%);Air插棒HU值比较均匀,其余密度插棒HU值随扫描条件不同而存在显著性差异(F=660.84,P=0);一年内不同扫描条件HU均值显示较好的一致性。结论 不同扫描条件下CBCT影像HU值一致性存在一定差异,但具有较好的长期稳定性,从而为CBCT影像直接用于自适应放疗中剂量计算奠定基础。

         

        Abstract: Objective To evaluate the cone beam CT (CBCT) image quality and its long-term stability for HUs under different scanning conditions. Methods With the OBI system, Catphan 504 phantom using the different modes was scanned through one year. Analysis of variance (ANOVA) was used to evaluate the CBCT images. Results The results of contrast resolution under different scanning conditions met the requirements of acceptance test procedure. Spatial contrast resolution(≥ 5 lp/cm) of half-fan scanning conditions were inferior to those (≥ 7 lp/cm) of full-fan mode. The accuracies of spatial linear distance and angle were within ±1%. HUs in air density plug were constant, but HUs of other plugs with different conditions had the significant difference (F=660.84, P=0). Under different conditions in a year, the mean HUs showed a good agreement. Conclusions HUs of CBCT image under different scanning conditions exist a little difference, but there is a good consistency for the long term stability analysis. It is helpful for CBCT images directly used for dose re-calculation in adaptive radiation therapy (ART).

         

      /