热脱附-气相色谱-质谱法测定医用分子筛制氧机中的挥发性有机物

      Determination of Volatile Organic Compounds in Medical Molecular Sieve Oxygen Concentrators by Thermal Desorption-Gas Chromatography-Mass Spectrometry

      • 摘要: 采用热脱附-气相色谱-质谱法测定医用分子筛制氧机产生的挥发性有机物(VOCs)的方法。将制氧机采用分支采样的方式以0.5 L/min的流速用Tenax GR/carbopack B吸附管采集气体。将吸附管进行解析并通过程序升温进样系统进样,用SH-I-624Sil MS柱进行色谱分离。采用本方法在医用分子筛制氧机共检出4种VOCs(BHT-Q、PTBP、BHT-quinol、 EHB)。4种VOCs在3~100 ng范围内线性关系良好(R2>0.99),当采样体积为20L时,方法检出限为0.003 9~0.022 2 μg/m3,4种VOCs加标回收率在95%~115%范围内,相对标准偏差小于5%(n=6)。该方法操作简单快捷、灵敏度高、准确度好,能够满足医用分子筛制氧机中VOCs的检测需求。

         

        Abstract: A method for determining volatile organic compounds (VOCs) emitted from medical molecular sieve oxygen concentrators was developed using thermal desorption-gas chromatography-mass spectrometry (TD-GC-MS). The oxygen concentrator gas was sampled at a flow rate of 0.5 L/min through a branched sampling system onto Tenax GR/Carbopack B adsorption tubes. The adsorbed compounds were desorbed and introduced via a programmed temperature vaporization inlet system, followed by chromatographic separation on an SH-I-624Sil MS column. Four VOCs (BHT-Q, PTBP, BHT-quinol, and EHB) were detected in the medical oxygen concentrator using this method. Calibration curves for these compounds exhibited excellent linearity (R²>0.99) within the range of 3~100 ng. With a sampling volume of 20 L, the detection limit of the four VOCs ranged from 0.003 9 to 0.022 2 μg/m³. Spike recovery rates for the four VOCs were between 95% and 115%, with relative standard deviations (RSDs) below 5% (n=6). The method is simple, rapid, highly sensitive, and accurate, making it suitable for VOC detection in medical molecular sieve oxygen concentrators.

         

      /