Abstract:
Objective To establish a high-performance liquid chromatography(HPLC) method for acetate of dialysis fluid and concentrates for hemodialysis.
Methods The column used is Synergi Polar-RP. Phosphate Buffered Saline(50 mmol/L, pH=2.5)is used as mobile phase. The flow rate is 1.0 mL/min. The wavelength of detection is 212 nm.
Results The linear relationship of acetate is good in the range of 0.1~20 mmol/L, r=0.9999. The recoveries were from 98.9%~99.5% RSD<0.5% (n=3).
Conclusion This method can easily and accurately determine the content of acetate ions in dialysate and dialysate concentrates, and can be applied to quality control in the production and use of such products.