植入式神经刺激器电磁兼容性体模仿真研究

      Torso Simulation Research on Electromagnetic Compatibility of Implantable Neurostimulator

      • 摘要: 行业标准YY 0989.3—2023《手术植入物 有源植入式医疗器械 第3部分:植入式神经刺激器》中27.106条款针对电磁干扰的防护试验,仅简略规定了电磁场暴露的试验电平,对试验用体模的参数未作详细规定。该研究旨在通过建模仿真,探讨不同体模参数在电磁场暴露下的内部电场分布情况,探索电场分布的规律。研究结果显示,体模的参数对其内部电场分布具有显著影响。该研究为植入式神经刺激器类产品的电磁兼容性试验提供了重要依据和参考。

         

        Abstract: In the YY 0989.3—2023 standard, clause 27.106 specifies the protection test against electromagnetic interference, but it only briefly describes the test level for electromagnetic exposure, and does not detail the parameters of the torso. This study aims to explore the internal field distribution for different torso parameters under electromagnetic exposure, and explore the patterns of field distribution through modeling and simulation. The results indicate that the parameters of the torso significantly affect the internal field distribution. The findings of this study provide a basis and reference for the electromagnetic compatibility test for implantable neurostimulator products.

         

      /

      返回文章
      返回